Vet Med - Czech, 2021, 66(2):66-71 | DOI: 10.17221/39/2020-VETMED
Comparison of three ECG machines for electrocardiography in green iguanas (Iguana iguana)Original Paper
- 1 Avian and Exotic Animal Clinic, Faculty of Veterinary Medicine, University of Veterinary and Pharmaceutical Sciences Brno, Brno, Czech Republic
- 2 Private Veterinarian, Nové Veselí, Czech Republic
- 3 Clinique Vétérinaire du Village d'Auteuil, Paris, France
The aim of the study was to compare the heart rate, QRS interval, and R wave amplitude across three electrocardiogram models, and assess the ability of each of them to provide electrocardiograms (ECG) for clinical interpretation. The three electrocardiogram models included ECG Seiva Praktik Veterinary, CardioStore ECG and AliveCor Veterinary Heart Monitor. The data were collected from twelve healthy adult captive green iguanas (Iguana iguana) monitored under a manual restraint at a room temperature of 22.6-28.0 °C. The ECGs using the Seiva Praktik and CardioStore ECG veterinary electrocardiography were performed with standard 4 lead ECG recordings. The AliveCor Veterinary Heart Monitor was placed (with the use of gel) directly on the lateral body wall. The mean heart rate was 42 ± 8 beats/min (CardioStore), 50 ± 11 beats/min (Seiva Praktik Veterinary), and 51 ± 9 beats/min (AliveCor Veterinary Heart Monitor). No significant difference in the heart rate was observed. A significant difference (P < 0.05) in the QRS duration was observed between the CardioStore and AliveCor Veterinary Heart Monitor. Significant differences (P < 0.01) in the R wave amplitude were detected between the CardioStore and AliveCor Veterinary Heart Monitor and between the Seiva Praktik Veterinary and AliveCor Veterinary Heart Monitor. The ECGs produced by the Seiva Praktik VVeeterinary and CardioStore machines were interpretable at 100%, while those produced by the AliveCor Veterinary Heart Monitor were interpretable at 66%. Seiva Praktik Veterinary is most appropriately used as an anaesthesia monitoring tool. AliveCor Veterinary Heart Monitor could be used as an additional diagnostic tool, but the results should be ideally confirmed with a standard ECG machine. Seiva Praktik Veterinary is the most appropriate tool for monitoring the ECG within the anaesthesia, while CardioStore might be most appropriately used as an advanced diagnostic tool by virtue of its software assistance. The ECGs obtained with AliveCor Veterinary Heart Monitor should be confirmed using a standard ECG machine.
Keywords: reptile cardiology; reptile electrocardiogram; heart frequency; R wave; QRS complex
Received: February 13, 2020; Accepted: November 10, 2020; Published: February 28, 2021 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- Bado O, Dlouha M, Kolmanova E, Frydrych M. Influence of new ultra-short-acting β-blockers on selected physiological indicators in laboratory rats. Vet Med-Czech. 2017 Sep;62(9):493-507.
Go to original source...
- Bogan JE. Ophidian cardiology - A review. J Herpetol Med Surg. 2017;27(1-2):62-77.
Go to original source...
- Boukens BJD, Kristensen DL, Filogonio R, Carreira LBT, Sartori MR, Abe AS, Currie S, Joyce W, Conner J, Opthof T, Crossley DA 2nd, Wang T, Jensen B. The electrocardiogram of vertebrates: Evolutionary changes from ectothermy to endothermy. Prog Biophys Mol Biol. 2019 Jul;144:16-29.
Go to original source...
Go to PubMed...
- Dahhan M. Elektrokardiographische Untersuchungen beim Grunen Leguan (Iguana iguana) [Electrocardiographic parameters in the green iguana (Iguana iguana)] [dissertation]. München: Ludwig-Maximilians-Universität; 2006. 101 p. German.
- Davies F, Francis ETB, King TS. Electrocardiogram of the crocodilian heart. Nature. 1951 Jan 27;167(4239):146.
Go to original source...
Go to PubMed...
- Germer CM, Tomaz JM, Carvalho AF, Bassani RA, Bassani JWM. Electrocardiogram, heart movement and heart rate in the awake gecko (Hemidactylus mabouia). J Comp Physiol. 2015 Jan;185(1):111-8.
Go to original source...
Go to PubMed...
- Haberman ZC, Jahn RT, Bose R, Tun H, Shinbane JS, Doshi RN, Chang PM, Saxon LA. Wireless smartphone ECG enables large-scale screening in diverse populations. J Cardiovasc Electrophysiol. 2015 May;26(5):520-6.
Go to original source...
Go to PubMed...
- Heaton-Jones T, King R. Characterization of the electrocardiogram of the American alligator (Alligator mississippiensis). J Zoo Wildl Med. 1994;25(1):40-7.
- Holz RM, Holz P. Electrocardiography in anesthetized redeared sliders (Trachemys scripta elegans). Res Vet Sci. 1995 Jan;58(1):67-9.
Go to original source...
Go to PubMed...
- Hunt CJ. Electrocardiography of the normal inland bearded dragon (Pogona vitticeps) [thesis]. [United Kingdom]: Royal College of Veterinary Surgeons; 2013. 43 p.
- Chung EH, Guise KD. QTC intervals can be assessed with the AliveCor heart monitor in patients on dofetilide for atrial fibrillation. J Electrocardiol. 2015 Jan-Feb;48(1):8-9.
Go to original source...
Go to PubMed...
- Jacob JS, McDonald HS. Temperature preferences and electrocardiography of Elaphe obsoleta (Serpentes). Comp Biochem Physiol Part A Physiol. 1975 Dec 1;52(4):591-4.
Go to original source...
Go to PubMed...
- Liu CH, Li R. Electrocardiogram and heart rate in response to temperature acclimation in three representative vertebrates. Comp Biochem Physiol. 2005 Dec;142(4):416-21.
Go to original source...
Go to PubMed...
- McDonald HS, Heath JE. Electrocardiographic observation on the tuatara, Sphenodon punctatus. Comp biochem Physiol. 1971 Dec;40(4):881-92.
Go to original source...
Go to PubMed...
- Mitchell M. Reptile cardiology. Clin North Am Exot Anim Pract. 2009 Jan;12(1):65-79.
Go to original source...
Go to PubMed...
- Mullen RK. Comparative electrocardiography of the Squamata. Physiol Zool. 1967;40(2):114-26.
Go to original source...
- Schilliger L, Girling S. Cardiology. In: Divers S, Stahl SJ, editors. Mader's reptile and amphibian medicine and surgery. 3rd ed. St. Louis: Elsevier; 2019. p. 669-98.
Go to original source...
- Schilliger L, Chai N, Chetboul V, Bonwitt J. Smartphonebased ECG monitor: Feasibility and applications in herpetological medicine [abstract]. In: Proceedings Association of Reptilian and Amphibian Veterinarians; 2014 Oct 18-24; Orlando, Florida. Orlando, FL,USA: Arav; 2014. p. 129-30.
- Schumacher J, Mans CH. Anesthesia. In: Divers SJ, Mader DR, editors. Current therapy in reptile medicine and surgery. St. Louis: Saunders Elsevier; 2014. p. 134-53.
Go to original source...
- Smith J, Ward J, Urbano T, Mueller M. Use of AliveCor heart monitor for heart rate and rhythm evaluation in dairy water buffalo calves (Bubalis Bubalis). J Dai Vet Anim Res. 2016;4(2):261-4.
Go to original source...
- Valentinuzzi ME, Hoff HE, Geddes LA. Electrocardiogram of the snake: Intervals and durations. J Electrocardiol. 1969;2(4):343-52.
Go to original source...
Go to PubMed...
- Valentinuzzi ME, Hoff HE, Geddes LA. Electrocardiogram of the snake: Effect of vagal stimulation on the Q-T duration. J Electrocardiol. 1970;3(1):21-7.
Go to original source...
Go to PubMed...
- Vandenberk T, Stans J, Mortelmans C, Van Haelst R, Van Schelvergem G, Pelckmans C, Smeets CJ, Lanssens D, De Canniere H, Storms V, Thijs IM, Vaes B, Vandervoort PM. Clinical validation of heart rate apps: Mixed-methods evaluation study. JMIR Mhealth Uhealth. 2017 Aug 25; 5(8): [15].
Go to original source...
Go to PubMed...
- Vezzosi T, Buralli C, Marchesotti F, Porporato F, Tognetti R, Zini E, Domenech O. Diagnostic accuracy of a smartphone electrocardiography in dogs: Comparison with standard 6-lead electrocardiography. Vet J. 2016 Oct;216:33-7.
Go to original source...
Go to PubMed...
- Zemanova A, Sochorcova V, Knotek Z. Zakladni metody kardiologickeho vysetreni plazu [Basic methods of cardiological examination in reptiles]. Veterinarni klinika. 2016;13:20-7. Czech.
This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International (CC BY NC 4.0), which permits non-comercial use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.