Vet Med - Czech, 2017, 62(5):285-291 | DOI: 10.17221/25/2017-VETMED
Comparison of continuous and intermittent glucose monitoring systems in a dog with diabetic ketoacidosis: a case reportCase Report
- 1 College of Veterinary Medicine, Chonnam National University, Gwangju, Republic of Korea
- 2 College of Veterinary Medicine, Chonbuk National University, Iksan, Republic of Korea
A Miniature schnauzer (12 years old, neutered male) was referred for lethargy, anorexia, and oral bleeding. On initial evaluation, severe hyperglycaemia (blood glucose concentration of 34.9 mmol/l), ketonuria, systemic inflammation (fever, panting, left-shift neutrophilia, and a high C-reactive protein level of 980.97 nmol/l, abnormal pancreatic lipase immunoreactivity, and periodontitis were found. With consideration of possible insulin resistance, blood glucose (BG) levels were monitored using a portable glucose meter (q 1-3 h) and a continuous glucose monitoring system (CGMS) for 72 h (three consecutive trials); intensive insulin therapy was initiated using regular insulin (2.2 IU/kg intravenously). The insulin doses needed, based on the nadir, peak, and duration of insulin action from a traditional intermittent glucose curve were higher than those based on the CGMS results. Meanwhile, transient hyperglycaemic and hypoglycaemic periods, occurring between the intermittent measurements, were easily identified with the CGMS. Therefore, insulin resistance and the Somogyi phenomenon are less likely to occur with use of the CGMS than with intermittent BG measurements. By comparing data from a CGMS to those from an intermittent portable BG measurement system, this case report emphasises the importance and usefulness of a CGMS.
Keywords: continuous glucose monitoring system; portable glucometer; diabetic ketoacidosis; hyperglycaemia; hypoglycaemia
Published: May 31, 2017 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- Affenzeller N, Thalhammer JG, Willmann M (2011): Homebased subcutaneous continuous glucose monitoring in 10 diabetic dogs. Veterinary Record 169, 206.
Go to original source...
Go to PubMed...
- Brunner R, Kitzberger R, Miehsler W, Herkner H, Madl C, Holzinger U (2011): Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients. Critical Care Medicine 39, 659-664.
Go to original source...
Go to PubMed...
- Choudhary P, Ramasamy S, Green L, Gallen G, Pender S, Brackenridge A, Amiel SA, Pickup JC (2013): Real-time continuous glucose monitoring significantly reduces severe hypoglycemia in hypoglycemia-unaware patients with type 1 diabetes. Diabetes Care 36, 4160-4162.
Go to original source...
Go to PubMed...
- Collings AJ, Gow AG, Marques A, Yool D, Furneaux R, Mellanby R, Watson PJ (2012): A prospective study of basal insulin concentrations in dogs with congenital portosystemic shunts. Journal of Small Animal Practice 53, 228-233.
Go to original source...
Go to PubMed...
- Corradini S, Pilosio B, Dondi F, Linari G, Testa S, Brugnoli F, Gianella P, Pietra M, Fracassi F (2016): Accuracy of a flash glucose monitoring system in diabetic dogs. Journal of Veterinary Internal Medicine 30, 983-988.
Go to original source...
Go to PubMed...
- Corstjens AM, Ligtenberg JJM, van der Horst ICV, Spanjersberg R, Lind JSW, Tulleken JE, Meertens JHJM, Zijlstra JG (2006): Accuracy and feasibility of point-of-care and continuous blood glucose analysis in critically ill ICU patients. Critical Care 10, R135.
Go to original source...
Go to PubMed...
- DeClue AE, Cohn LA, Kerl ME, Wiedmeyer CE (2004): Use of continuous blood glucose monitoring for animals with diabetes mellitus. Journal of the American Animal Hospital Association 40, 171-173.
Go to original source...
Go to PubMed...
- Dunayer EK (2004): Hypoglycemia following canine ingestion of xylitol-containing gum. Veterinary and Human Toxicology 46, 87-88.
- Garg S, Zisser H, Schwartz S, Bailey T, Kaplan R, Ellis S, Jovanovic L (2006): Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: A randomized controlled trial. Diabetes Care 29, 44-50.
Go to original source...
Go to PubMed...
- Harris DL, Battin MR, Weston PJ, Harding JE (2010): Continuous glucose monitoring in newborn babies at risk of hypoglycemia. The Journal of Pediatrics 157, 198-202.
Go to original source...
Go to PubMed...
- Hess RS (2010): Insulin resistance in dogs. Veterinary Clinics of North America. Small Animal Practice 40, 309-316.
Go to original source...
Go to PubMed...
- Hoedemaekers CWE, Gunnewiek JMTK, Prinsen MA, Willems JL, Hoeven JGD (2008): Accuracy of bedside glucose measurement from three glucometers in critically ill patients. Critical Care Medicine 36, 3062-3066.
Go to original source...
Go to PubMed...
- Holzinger U, Warszawska J, Kitzberger R, Herkner H, Metnitz PGH, Madl C (2009): Impact of shock requiring norepinephrine on the accuracy and reliability of subcutaneous continuous glucose monitoring. Intensive Care Medicine 35, 1383-1389.
Go to original source...
Go to PubMed...
- Hume DZ, Drobatz KJ, Hess RS (2006): Outcome of dogs with diabetic ketoacidosis: 127 dogs (1993-2003). Journal of Veterinary Internal Medicine 20, 547-555.
Go to original source...
- Joubert M, Reznik Y (2012): Personal continuous glucose monitoring (CGM) in diabetes management: Review of the literature and implementation for practical use. Diabetes Research and Clinical Practice 96, 294-305.
Go to original source...
Go to PubMed...
- Loose NL, Rudloff E, Kirby R (2008): Hypoglycemia and its effect on the brain. Journal of Veterinary Emergency and Critical Care 18, 223-234.
Go to original source...
- Miller SI, Wallace RJ, Musher DM, Septimus EJ, Kohl S, Baughn RE (1980): Hypoglycemia as a manifestation of sepsis. The American Journal of Medicine 68, 649-654.
Go to original source...
Go to PubMed...
- Nelson RW, Reusch CE (2014): Animal models of disease: classification and etiology of diabetes in dogs and cats. Journal of Endocrinology 222, T1-T9.
Go to original source...
Go to PubMed...
- Prabhudesai S, Kanjani A, Bhagat I, Ravikumar KG, Ramachandran B (2015): Accuracy of a real-time continuous glucose monitoring system in children with septic shock: A pilot study. Indian Journal of Critical Care Medicine 19, 642-647.
Go to original source...
Go to PubMed...
- Reineke EL, Fletcher DJ, King LG, Drobatz KJ (2010): Accuracy of a continuous glucose monitoring system in dogs and cats with diabetic ketoacidosis. Journal of Veterinary Emergency and Critical Care 20, 303-312.
Go to original source...
Go to PubMed...
- Surman S, Fleeman L (2013): Continuous glucose monitoring in small animals. Veterinary Clinics of North America: Small Animal Practice 43, 381-406.
Go to original source...
Go to PubMed...
- The NICE-SUGAR Study Investigators (2012): Hypoglycemia and risk of death in critically ill patients. New England Journal of Medicine 367, 1108-1118.
Go to original source...
Go to PubMed...
- Wiedmeyer CE, DeClue AE (2008): Continuous glucose monitoring in dogs and cats. Journal of Veterinary Internal Medicine 22, 2-8.
Go to original source...
Go to PubMed...
- Zini E, Glaus TM, Minuto F, Arvigo M, Hauser B, Reusch CE (2007): Paraneoplastic hypoglycemia due to an insulin-like growth factor type-II secreting hepatocellular carcinoma in a dog. Journal of Veterinary Internal Medicine 21, 193-195.
Go to original source...
This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International (CC BY NC 4.0), which permits non-comercial use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.